Specifications – CS-1000A Spectroradiometer

Model

CS-1000A SPECTRORADIOMETER

Wavelength range

380 to 780 nm

Wavelength resolution

0.9nm/pixel

Spectral bandwidth

5nm

Acceptance angle

1o

Minimum measuring area

7.9mm (with standard lens),  1.15mm (with macro lens)

Minimum measuring distance

45cm (with standard lens), 20cm (with macro lens)

Luminance display range

0.01 to 80,000 cd/m2 (for Illuminant A)

Accuracy (for Illuminant A)

± 2% ± 1 digit, x: ± 0.0015, y: ± 0.001

Luminance range:

1 to 8,000 cd/m2 with standard lens,

10 to 80,000 cd/m2 with macro lens

300 to 800,000 cd/m2 with CS-A6

3,000 to 8,000,000 cd/m2 with CS-A7)

Repeatability

Luminance: 0.1% + 1 digit, Chromaticity: x: 0.0002, y:0.0002

Luminance range:

3 to 8,000 cd/m2 with standard lens,

30 to 80,000 cd/m2 with macro lens

300 to 800,000 cd/m2 with CS-A6

3,000 to 8,000,000 cd/m2 with CS-A7)

Spectral accuracy

± 0.3nm (Hg lamp/Mean wavelength)

Polarization error

Less than 5% (400nm to 780nm)

Integration time

40 msec to 60 sec

Display

Le, Lvxy, Lv u' v', LvTDuv, DLvxy, DLvu'v', (Observer can be switched between 2o and 10o)

Data Memory

Measurement data: 30 sets, Target data: 20 sets

Interface

RS-232C

Power

AC100V/120V/230V 50/60Hz (using AC adapter AC-A12)

Operating environment conditions

Temperature: 5 to 35oC (41 to 95oF); relative humidity 85% or less (at 35oC/95oF) with no condensation.

Maximum altitude: 2000m

Installation category: I

Pollution degree: 2

Storage temperature range

0 to 45oC (32 to 113oF); relative humidity 85% or less (at 35oC/95oF) with no condensation

Dimensions (W x H x D)

146 x 148 x 256mm (5-3/4 x 5-13/16 x 10-1/16 in.)

Weight

4.7kg (10.38 lb)

Optional accessories

 

 

ND Filter CS-A6 (1/10),CS-A7 (1/100):

High-luminance light sources which exceed the possible measurement range of the CS-1000 can also be measured by connecting the measuring instrument to the CS-1000

White Calibration Plate, CS-A5:

This unit can be used for white calibration when reflective substance colors and transparent substance colors are measured. The calibration data disks are packaged with the white calibration board.

Panhead CS-A4/Tripod CS-A3