Model |
CS-200 |
||||||||||||||||||
Measurement |
|
||||||||||||||||||
Accuracy |
|
||||||||||||||||||
Repeatability |
(Repeatability in superFAST & FAST mode is SLOW mode multiplied by 2) |
||||||||||||||||||
Measurement |
0.5 sec / meas. (superFAST) 1 sec / meas. (FAST) |
||||||||||||||||||
Measurement |
Spectral method, Grating + linear photo diode array |
||||||||||||||||||
Measuring |
0.1° , 0.2°, 1° (selectable) |
||||||||||||||||||
Minimum |
0.5 mm (standard) 0.1 mm(close up lens) |
||||||||||||||||||
Minimum |
296 mm - Distance from front edge of metal lens barrel. |
||||||||||||||||||
Observer |
2° and 10° |
||||||||||||||||||
Color Space |
Lv x y, Lv u' v', Lv T Δuv, XYZ, dominant wavelength |
||||||||||||||||||
Measurement Sychronization |
Vertical Synchronization frequency: 40.00 to 200.00 Hz |
||||||||||||||||||
Interface |
USB 1.1 |
||||||||||||||||||
Power Source |
AC-Adapter / 4 AA size battery |
||||||||||||||||||
Battery life |
Approx. 3 hours (continuous measurement/ Fast mode/ AA size Alkaline cells) |
||||||||||||||||||
Size |
95 mm (W) x 127 mm (H) x 330 mm (L) |
||||||||||||||||||
Weight |
1.8 kg (without battery) |
||||||||||||||||||
Operating |
0-40°C / less than 85%RH (at 35°C) no condensation |
||||||||||||||||||
Storage |
0-45°C / less than 85%RH (at 35°C) no condensation |
||||||||||||||||||
Standard |
AC-Adapter, lens cap, switch cover, ND-eyepiece filter, USB-cable (2m), Data processing software CS-S10W (Standard version) |
||||||||||||||||||
Optional |
Soft case, close up lenses(No. 107 & No. 122), step up ring (40.5-55mm), ND-filter (1/10 & 1/100), angle finder, White calibration plate (45-0) & (d-0), Data processing software CS-S10W (Professional version) |
||||||||||||||||||
# 1 |
23°C +/-2°C |
Lv = 0.01-10 cd/m² |
SLOW |
average of 30 measurements |
|
|
Lv = 10 cd/m² and higher |
SLOW |
average of 10 measurements |
#2 At 0.2° measuring angle, the amount of received light is approx. 1/25 of that for
1°. Therefore, the repeatability becomes the same as that for 1° with 25 times smaller luminance.
At 0.1° measuring angle, the amount of received light is approx. 1/100 of that for 1°. Therefore, the repeatability becomes the same as that for 1° with 100 times smaller luminance.
Specifications are subject to change without notice.